Validation of ammonia diffusive and pumped samplers in a controlled atmosphere test facility using traceable Primary Standard Gas Mixtures
Martin, Nicholas A.; Ferracci, Valerio; Cassidy, Nathan; Hook, Josh; Battersby, Ross M.; di Meane, Elena Amico; Tang, Yuk S.; Stephens, Amy C.M.; Leeson, Sarah R.; Jones, Matthew R.; Braban, Christine F. ORCID: https://orcid.org/0000-0003-4275-0152; Gates, Linda; Hangartner, Marcus; Stoll, Jean-Marc; Sacco, Paolo; Pagani, Diego; Hoffnagle, John A.; Seitler, Eva. 2019 Validation of ammonia diffusive and pumped samplers in a controlled atmosphere test facility using traceable Primary Standard Gas Mixtures. Atmospheric Environment, 199. 453-462. 10.1016/j.atmosenv.2018.11.038
Before downloading, please read NORA policies.Preview |
Text
N521795PP.pdf - Accepted Version Available under License Creative Commons Attribution Non-commercial No Derivatives 4.0. Download (3MB) | Preview |
Abstract/Summary
We report the determination of ammonia (NH3) diffusive sampling rates for six different designs of commercial diffusive samplers (CEH ALPHA sampler, Gradko diffusion tube, Gradko DIFRAM-400, Passam ammonia sampler,and ICS Maugeri Radiello radial sampler (blue and white turbulence barriers)), together with the validation test results for a pumped sampler (CEH DELTA denuder). The devices were all exposed in the UK's National Physical Laboratory's (NPL) controlled atmosphere test facility (CATFAC). For each of the seven diffusive sampler exposure tests there were traceable concentrations of ammonia (in the range 3–25 μgm−3) generated under well-defined conditions of temperature, relative humidity and wind speed, which are applicable to a variety of ambient monitoring environments. The sampler exposure time at each concentration was 28 days, except for the radial devices, which were exposed for 14 days. The work relied on the dilution of newly developed stable Primary Standard Gas Mixtures (PSMs) prepared by gravimetry in passivated gas cylinders as a method of improving the metrological traceability of ammonia measurements. The exposed diffusive samplers were sent blind to the participants for analysis and the reported NH3 concentrations were then compared against the known reference concentration. From the results for each sampler type a diffusive sampling rate was calculated and compared against the rate used routinely by the participants. Some measurement results were in good agreement with the known traceable reference concentration (particularly for one diffusive sampler design (ALPHA)), while other devices exhibited over-reading and under-reading (each with a clear bias). The new diffusive sampling rates determined in the laboratory study were then applied to measurements in a field comparison campaign, and this was found to deliver an improvement in agreement between the different devices deployed.
Item Type: | Publication - Article |
---|---|
Digital Object Identifier (DOI): | 10.1016/j.atmosenv.2018.11.038 |
UKCEH and CEH Sections/Science Areas: | Atmospheric Chemistry and Effects (Science Area 2017-) |
ISSN: | 1352-2310 |
Additional Keywords: | ammonia, diffusive samplers, ALPHA samplers, DELTA, validation, diffusive sampling rates, denuders, metrological traceability |
NORA Subject Terms: | Atmospheric Sciences |
Date made live: | 07 Dec 2018 14:59 +0 (UTC) |
URI: | https://nora.nerc.ac.uk/id/eprint/521795 |
Actions (login required)
View Item |
Document Downloads
Downloads for past 30 days