Design, simulation and characterisation of integrated optics for a microfabricated flow cytometer
Barat, David; Benazzi, Guiseppe; Mowlem, Matthew Charles; Ruano, Jesus Miguel; Morgan, Hywel. 2010 Design, simulation and characterisation of integrated optics for a microfabricated flow cytometer. Optics Communications, 283 (9). 1987-1992. 10.1016/j.optcom.2009.12.046
Full text not available from this repository. (Request a copy)Abstract/Summary
Flow cytometry is widely used for analyzing micro-particles such as cells and bacteria. Microfabricated flow cytometers promise reduced instrument size and cost with increased robustness and have application in medicine, life sciences and environmental metrology. Further miniaturisation and robustness can be achieved if integrated optics are used instead of traditional free space optics. We present designs simulation and experimental characterisation of integrated optics for a microfabricated cytometer made from SU-8 resin on a glass substrate. The optics constructed from combinations of optical fibres (positioned with microgrooves), waveguides, and microlenses enable analysis of scattered light and fluorescence from particles positioned near the centre of a microchannel using one dimensional sheath flow. Four different methods for directing the incident light onto the particles are examined and the optimum design discussed.
Item Type: | Publication - Article |
---|---|
Digital Object Identifier (DOI): | 10.1016/j.optcom.2009.12.046 |
ISSN: | 0030-4018 |
Date made live: | 26 Jan 2010 +0 (UTC) |
URI: | https://nora.nerc.ac.uk/id/eprint/172160 |
Actions (login required)
View Item |
Document Downloads
Downloads for past 30 days
Downloads per month over past year