Practical field calibration of portable monitors for mobile measurements of multiple air pollutants

Lin, Chun; Masey, Nicola; Wu, Hao; Jackson, Mark; Carruthers, David J.; Reis, Stefan ORCID: https://orcid.org/0000-0003-2428-8320; Doherty, Ruth M.; Beverland, Iain; Heal, Mathew R.. 2017 Practical field calibration of portable monitors for mobile measurements of multiple air pollutants. Atmosphere, 8 (12), 231. 19, pp. https://doi.org/10.3390/atmos8120231

Before downloading, please read NORA policies.
[img]
Preview
Text
N518508JA.pdf - Published Version
Available under License Creative Commons Attribution 4.0.

Download (948kB) | Preview

Abstract/Summary

To reduce inaccuracies in the measurement of air pollutants by portable monitors it is necessary to establish quantitative calibration relationships against their respective reference analyser. This is usually done under controlled laboratory conditions or one-off static co-location alongside a reference analyser in the field, neither of which may adequately represent the extended use of portable monitors in exposure assessment research. To address this, we investigated ways of establishing and evaluating portable monitor calibration relationships from repeated intermittent deployment cycles over an extended period involving stationary deployment at a reference site, mobile monitoring, and completely switched off. We evaluated four types of portable monitors: Aeroqual Ltd. (Auckland, New Zealand) S500 O3 metal oxide and S500 NO2 electrochemical; RTI (Berkeley, CA, USA) MicroPEM PM2.5; and, AethLabs (San Francisco, CA, USA) AE51 black carbon (BC). Innovations in our study included: (i) comparison of calibrations derived from the individual co-locations of a portable monitor against its reference analyser or from all the co-location periods combined into a single dataset; and, (ii) evaluation of calibrated monitor estimates during transient measurements with the portable monitor close to its reference analyser at separate times from the stationary co-location calibration periods. Within the ~7 month duration of the study, ‘combined’ calibration relationships for O3, PM2.5, and BC monitors from all co-locations agreed more closely on average with reference measurements than ‘individual’ calibration relationships from co-location deployment nearest in time to transient deployment periods. ‘Individual’ calibrations relationships were sometimes substantially unrepresentative of the ‘combined’ relationships. Reduced quantitative consistency in field calibration relationships for the PM2.5 monitors may have resulted from generally low PM2.5 concentrations that were encountered in this study. Aeroqual NO2 monitors were sensitive to both NO2 and O3 and unresolved biases. Overall, however, we observed that with the ‘combined’ approach, ‘indicative’ measurement accuracy (±30% for O3, and ±50% for BC and PM2.5) for 1 h time averaging could be maintained over the 7-month period for the monitors evaluated here.

Item Type: Publication - Article
Digital Object Identifier (DOI): https://doi.org/10.3390/atmos8120231
UKCEH and CEH Sections/Science Areas: Atmospheric Chemistry and Effects (Science Area 2017-)
ISSN: 2073-4433
Additional Information. Not used in RCUK Gateway to Research.: Open Access paper - full text available via Official URL link.
Additional Keywords: air pollution sensor, air quality, O3, NO2, PM2.5, black carbon, personal exposure
NORA Subject Terms: Atmospheric Sciences
Date made live: 27 Nov 2017 11:13 +0 (UTC)