Note: a dual temperature closed loop batch reactor for determining the partitioning of trace gases within CO2-water systems
Warr, Oliver; Rochelle, Christopher A.; Masters, Andrew J.; Ballentine, Christopher J.. 2016 Note: a dual temperature closed loop batch reactor for determining the partitioning of trace gases within CO2-water systems. Review of Scientific Instruments, 87 (1), 016102. https://doi.org/10.1063/1.4938159
Before downloading, please read NORA policies.
|
Text (Open Access Paper)
1.4938159.pdf - Published Version Available under License Creative Commons Attribution 4.0. Download (1MB) | Preview |
Abstract/Summary
An experimental approach is presented which can be used to determine partitioning of trace gases within CO2-water systems. The key advantages of this system are (1) The system can be isolated with no external exchange, making it ideal for experiments with conservative tracers. (2) Both phases can be sampled concurrently to give an accurate composition at each phase at any given time. (3) Use of a lower temperature flow loop outside of the reactor removes contamination and facilitates sampling. (4) Rapid equilibration at given pressure/temperature conditions is significantly aided by stirring and circulating the water phase using a magnetic stirrer and high-pressureliquid chromatography pump, respectively.