nerc.ac.uk

Characterisation and deployment of an immobilised pH sensor spot towards surface ocean pH measurements

Clarke, Jennifer S.; Achterberg, Eric P.; Rérolle, Victoire M.C.; Abi Kaed Bey, Samer; Floquet, Cedric F.A.; Mowlem, Matthew C.. 2015 Characterisation and deployment of an immobilised pH sensor spot towards surface ocean pH measurements. Analytica Chimica Acta, 897. 69-80. https://doi.org/10.1016/j.aca.2015.09.026

Before downloading, please read NORA policies.
[img]
Preview
Text (Open Access paper)
1-s2.0-S0003267015011654-main.pdf - Published Version
Available under License Creative Commons Attribution 4.0.

Download (1MB) | Preview

Abstract/Summary

The oceans are a major sink for anthropogenic atmospheric carbon dioxide, and the uptake causes changes to the marine carbonate system and has wide ranging effects on flora and fauna. It is crucial to develop analytical systems that allow us to follow the increase in oceanic pCO2 and corresponding reduction in pH. Miniaturised sensor systems using immobilised fluorescence indicator spots are attractive for this purpose because of their simple design and low power requirements. The technology is increasingly used for oceanic dissolved oxygen measurements. We present a detailed method on the use of immobilised fluorescence indicator spots to determine pH in ocean waters across the pH range 7.6–8.2. We characterised temperature (−0.046 pH/°C from 5 to 25 °C) and salinity dependences (−0.01 pH/psu over 5–35), and performed a preliminary investigation into the influence of chlorophyll on the pH measurement. The apparent pKa of the sensor spots was 6.93 at 20 °C. A drift of 0.00014 R (ca. 0.0004 pH, at 25 °C, salinity 35) was observed over a 3 day period in a laboratory based drift experiment. We achieved a precision of 0.0074 pH units, and observed a drift of 0.06 pH units during a test deployment of 5 week duration in the Southern Ocean as an underway surface ocean sensor, which was corrected for using certified reference materials. The temperature and salinity dependences were accounted for with the algorithm, R=0.00034−0.17·pH+0.15·S2+0.0067·T−0.0084·S·1.075R=0.00034−0.17·pH+0.15·S2+0.0067·T−0.0084·S·1.075. This study provides a first step towards a pH optode system suitable for autonomous deployment. The use of a short duration low power illumination (LED current 0.2 mA, 5 μs illumination time) improved the lifetime and precision of the spot. Further improvements to the pH indicator spot operations include regular application of certified reference materials for drift correction and cross-calibration against a spectrophotometric pH system. Desirable future developments should involve novel fluorescence spots with improved response time and apparent pKa values closer to the pH of surface ocean waters.

Item Type: Publication - Article
Digital Object Identifier (DOI): https://doi.org/10.1016/j.aca.2015.09.026
ISSN: 00032670
Additional Keywords: Seawater; pH sensor; Fluorescence; Optode; Ocean acidification
Date made live: 16 Nov 2015 09:40 +0 (UTC)
URI: https://nora.nerc.ac.uk/id/eprint/512231

Actions (login required)

View Item View Item

Document Downloads

Downloads for past 30 days

Downloads per month over past year

More statistics for this item...