nerc.ac.uk

An approach for correcting magnitude and phase distortion in wideband piezoelectric transducer systems

Assous, Siad; Gunn, David; Hopper, Clare; Jackson, Peter; Linnett, Laurie; Lovell, Mike. 2007 An approach for correcting magnitude and phase distortion in wideband piezoelectric transducer systems. In: Oceans 2007, Aberdeen, Scotland, 18-21 June 2007. 1543-1548.

Before downloading, please read NORA policies.
[img]
Preview
Text
Assous_Ocean2007_(5).pdf

Download (211kB)

Abstract/Summary

Abstract—Acoustic ultrasonic measurements are widespread and commonly performed using sensitive piezoelectric sensors. An accurate transducer system response to investigate pressure fluctuations in water and their subsequent detection remains a challenge. Typically, these sensors exploit the resonant behaviour of the piezoelectric active element, being designed to give maximum sensitivity in the bandwidth of interest. Calibration of such transducers can provide both magnitude and phase information describing the way in which the sensor responds to a surface displacement over its frequency range. Such resonant sensors are widely used for ultrasonic applications. The resonant nature of the sensors leads to the use of narrowband signals with central frequencies close to the resonant frequency of the piezoelectric element. Consequently, such devices work efficiently and linearly over only a narrow band of their overall frequency range. This causes phase and magnitude distortion of any linear broadband signal being transmitted through such a transmitter-receiver acoustic system. In the present work, we describe a software calibration technique to correct for distortion in a wideband piezoelectric transducer system. We consider only the input and the final output signals of the whole system. Compensating for the distortion of the magnitude and phase responses, we ensured the signal seen at the receiver represents a good replica of the desired signal. A Gaussian, linear, chirp signal was used to demonstrate our approach. This method may be applied to correct system distortion in a wide variety of ultrasonic applications.

Item Type: Publication - Conference Item (Paper)
Programmes: BGS Programmes > Seismology and Geomagnetism
Additional Keywords: Conference proceedings, acoustic, transducer, calibration
NORA Subject Terms: Earth Sciences
Date made live: 27 Jun 2008 15:01 +0 (UTC)
URI: https://nora.nerc.ac.uk/id/eprint/3423

Actions (login required)

View Item View Item

Document Downloads

Downloads for past 30 days

Downloads per month over past year

More statistics for this item...